Armstrong Steam QM-3 Steam Quality Monitor

Armstrong Steam QM-3 automatic steam quality monitor. Simultaneously measures steam dryness, superheat, and non-condensable gases (NCGs). EN 285 compliant with RS485/Modbus data logging.

Armstrong Steam QM-3 Steam Quality Monitor

The Armstrong Steam QM-3 is an automatic steam quality monitor intended to replace manual testing of pure steam and provide real-time data proving that steam quality meets applicable requirements. It performs three critical tests: calculating dryness, calculating superheat, and quantifying non-condensable gases (NCG). The NCG measurement is performed first; if it is within range, dryness and superheat measurements will be performed.

Armstrong Steam QM-3 Steam Quality Monitor

EN 285 Compliance

This European standard specifies requirements and the relevant tests for large steam sterilizers, primarily used for the sterilization of medical devices and their accessories but has been adopted by most pharmaceutical industries.

  • The sterilizer shall be designed to operate with saturated steam containing up to 3.5 ml non-condensable gases collected from 100 ml condensate.
  • The sterilizer shall be designed to operate with saturated steam with a dryness value not less than 0.95, where the dryness value denotes the mass of the gas fraction in the mass of saturated steam.
  • When the supplied steam is expanded to atmospheric pressure the superheat shall not exceed 25K.

The patented Steam Quality Monitor QM-3 is an alternative procedure to the one described in EN285 that has been shown to give equivalent results to the method specified in this European Standard.

Product Features

  • Simple “Plug and Play” installation.
  • Simultaneous steam dryness, superheat and non-condensables (NCGs) monitoring.
  • Safe alternative to the traditional manual method of sampling steam.
  • RS485 connection for data logging using regulation compliant device; results may be remotely monitored via MODBUS.

Manual Versus Automatic

Until now steam quality measurement has been a time-intensive, unreliable and potentially unsafe process. Steam QM-3 is not only more reliable and safer than manual testing, the unit is also portable, so it can be easily transported to multiple points on your steam line.

FeatureManual MethodAutomatic Method (Steam QM-3)
DescriptionA sample of clean steam is condensed, and the enthalpy allows measurement of steam dryness, and NCG content. Temperature measurement before condensation identifies an excess of superheat temperature.Reducing steam pressure to atmosphere allows measurement of steam dryness. Steam temperature and pressure measurements detect superheat. Volume of non-condensable gases are compared to condensate.
Disadvantages / AdvantagesTime Consuming: Typically requires two people, and can take up to three hours per measurement point.
Trending: Impossible to monitor a trend over time.
Unsafe: Inherent safety risks involved in sampling live steam.
Unreliable: Results depend on the skill of the technician.
Quick and Easy: Simple to install.
Trending: Continuous measurements provide trending data over time.
Safe: Installed while the steam valve is closed, making it much safer.
Reliable: Reliable and accurate within +/- 1% of steam dryness.

Specifications

ParameterSpecification
Calorimeter operating temperature range (T1)0 – 150 °C (32 – 302 °F)
Maximum allowable calorimeter temperature (T1)180 °C (356 °F)
Calorimeter operating saturated steam pressure range (P)0.5 – 4 barg (7 – 60 psig)
Superheated steam operating temperature range (T2)0 – 180 °C (32 – 356 °F)
Dryness fraction0.85 – 1
Dryness fraction accuracy± 0.01 of display
Non-condensable gases content≤ 15%
Condensate temperature range (T3)0 – 80 °C (32 – 176 °F)
Operating condensate temperature (T3)65 °C (149 °F)
Estimated steam consumption1.5 kg/h (3.3 lbs/h) @ 3 barg (45 psig)
Estimated water consumption15 L/h (4 gal/h) @ 10 °C (50 °F)
Electrical115/230 VAC (±10%), 50/60 Hz, 100 W

Dimensions & Weight

Dimensioninmm
A – Cabinet Width22550
B – Width24600
C – Cabinet Height401000
D – Depth7.5190
E – Width17430
F – Height37945
G - Height from inlet tee to bottom26660
H - Height from cabinet steam inlet to bottom7.5190
Cabinet Weight42 lb19 kg
Total Weight55 lb25 kg

Calorimeter Assembly & Cabinet Interior

Armstrong Steam QM-3 Calorimeter Assembly

Calorimeter Assembly Components:

  1. Inlet tee
  2. Sanitary clamp
  3. P pressure transmitter
  4. Calibrated orifice (inside connection)
  5. Calorimeter
  6. Sanitary clamp
  7. Steam line from calorimeter to superheating chamber (not visible)
  8. Superheating chamber
  9. Drain (6 mm)
  10. Tee
  11. T1 temperature transmitter
  12. T2 temperature transmitter
  13. Heating Element

Armstrong Steam QM-3 Cabinet Interior

Cabinet Interior Components:

  1. NCG pressure differential sensor
  2. EV1 solenoid valve
  3. NCG burette
  4. Condensate pressure differential sensor
  5. T3 temperature sensor
  6. Condensate burette
  7. EV3 solenoid valve
  8. EV2 solenoid valve
  9. EV0 solenoid valve
  10. Condenser
  11. Relay board with solenoid LED
  12. Main screen
  13. Cabinet steam inlet
  14. Cabinet condensate outlet

Installation Considerations

Burn Hazard

Uninsulated components upstream of cabinet may be hot. Do not touch when unit is working. Allow to cool before moving or servicing unit. Live steam will cause burns; condensate water may cause them. Skin exposure to 140 °F (60 °C) water for only five seconds may cause a second degree burn.

Shock Hazard

High voltages present inside equipment. Electrical installation must be performed by qualified personnel. Disconnect power before performing any electrical service.

  • Ambient temperature must be 5–60 °C (41–140 °F).
  • Relative humidity must be 30–80%.
  • Altitude must not exceed 2000 m (6562 ft).
  • Indoor use, Pollution Degree 2.
  • Attachment to vertical steam line is optimal. Attachment to horizontal steam line must be at center line or above. Connecting to top of horizontal line may result in elevated dryness readings.
  • Superheating chamber must be installed within 45° of vertical to front or back.

Armstrong Steam QM-3 Typical Installation for Vertical Steam Line

Armstrong Steam QM-3 Typical Installation for Horizontal Steam Line

Components and Parts List

DescriptionPart Nr.
Steam Quality Monitor QM-3 PackageD44212
Insulation jackets (2pcs)D43805
Stainless Steel Wall mountD44160
Calorimeter assemblyD49150
Condenser with fittingsD79386
Main board with displayD44124
Relay boardD44125
Vessel Non Condensable Gases assembly with expansion coilD44126
Vessel Condensate flow meter assembly with fittingsD46738
JUMO PT100 (T1 and T2)D44110
JUMO Pressure transmitterD44117
Heating elementD44118
BURKERT 3/2 solenoid valve (EV0)D44119
BURKERT 2/2 solenoid valve (EV1 and EV3)D44120
BURKERT 3/2 solenoid valve (EV2)D44121
JUMO PT100 (T3)D44122
JUMO Differential pressure sensorD44123
Orifice Plate with GasketD40020
Gasket PTFE for 1/2'' flexible hose (bag with 3 pieces)D53335

MODBUS Connection

A 9 ft (3 m) cord is provided with an M12 connector for Steam QM-3 cabinet connection. The logger end of the cord must be wired by the customer. Program the logger with the following information:

  • Baud Rate: 9600 bauds/sec
  • Data Length: 8 bits
  • Parity: None
  • Handshaking: N/A
  • Address: From 1 to 80
  • Mode: Master or slave

Master Mode

Data are sent in different registers using Modbus function 16 (0x10). All values are SI units. Decimal values are not used (e.g., 19.2 will show as 192).

RegisterNameFactorUnitData Type
40 001T1x 10°C(Integer value)
40 002T2x 10°C(Integer value)
40 004P1x 10bar[a](Integer value)
40 007Xx 1000N/A(Integer value)
40 008Qx 10kg/h(Integer value)
40 011% NCGx 10N/A(Integer value)
40 012AlarmsN/AN/ABit 1: Alarm 1, Bit 2: Alarm 2, Bit 3: Alarm 3, Bit 4: Default 4, Bit 5: Default 5, Bit 6: Default 6

Slave Mode

In slave mode, you can send queries with function 0x02 and 0x04. The length of the query and response must remain less than 23 bytes (max 3 registers per query).

Function 0x02 Read Discrete Inputs:

RegisterNameFactorUnitDescription
10001Alarm1N/A0: disable - 1: enableToggle to 1 when the alarm 1 is on
10002Alarm2N/A0: disable - 1: enableToggle to 1 when the alarm 2 is on
10003Alarm3N/A0: disable - 1: enableToggle to 1 when the alarm 3 is on
10004Default4N/A0: disable - 1: enableToggle to 1 when the default 4 is on
10005Default5N/A0: disable - 1: enableToggle to 1 when the default 5 is on
10006Default6N/A0: disable - 1: enableToggle to 1 when the default 6 is on

Function 0x04 Read Inputs Registers:

RegisterNameFactorUnitDescription
30001T110x°CInteger (16 bits)
30002T210x°CInteger (16 bits)
30004Alarm310xBar [a]Integer (16 bits)
30007Default41000x-Integer (16 bits)
30008Default510xKg/hrInteger (16 bits)
30011Default610x-Integer (16 bits)